– Press the yellow button of the handpiece. The device is activated and an acoustic signal sounds. At
the same time, the symbol in the field marked yellow is lit up yellow.
– Release the yellow button of the handpiece. The device is deactivated, and both the acoustic signal
and the yellow illumination go off.
– Press the blue button of the handpiece. The device is activated, an acoustic signal sounds, and at the
same time the symbol in the field marked blue is lit up blue.
– Activate the yellow button of the handpiece. The device remains activated in Coag position.
– Release the blue button of the handpiece. The device switches to Cut operating mode (audio signal
and illumination change).
– Release the yellow button of the handpiece. The device is deactivated, and both the acoustic signal
and the yellow illumination go off.
4.3.13 Activation using the (monopolar) ERBE-type
handpiece
– Connect the monopolar handpiece 265200 43 with cable 265200 45 to the monopolar HF socket.
– Connect the neutral electrode.
– Single-use double-faced electrodes should be connected for a suitably short time. The NE indicator on
the device should be lit up green.
– Press the yellow button of the handpiece. The device is activated, an acoustic signal sounds, and at
the same time the symbol in the field marked yellow is lit up yellow.
– Release the yellow button of the handpiece. The device is deactivated, and both the acoustic signal
and the yellow illumination go off.
– Press the blue button of the handpiece. The device is activated, an acoustic signal sounds, and at the
same time the symbol in the field marked blue is lit up blue.
– Also activate the yellow button of the handpiece; the device will remain activated in the Coag position.
– Release the blue button of the handpiece. The device switches to Cut operating mode (audio signal
and illumination change).
– Release the yellow button of the handpiece. The device is deactivated, and both the acoustic signal
and the yellow illumination go off.
4.3.14 Instrument detection with the MF socket
Note:
Since KARL STORZ produces only one multi-function (MF) instrument, there is no need to
check for any other. For ERBE instruments, which also fit, an error message should appear if
an adapter is used for simulation.
Note:
This test item is not applicable to device versions -111 or -112 (but does also apply to U111
and U112 versions). The detection of 26176LEB(L) cable for the multi-function socket is
implemented in versions –113 to –116 and U113 to U116.
When the cable is plugged into the MF socket, the following message will appear: “A-95: Instrument
recognized at MF socket”. If the cable is removed from the MF socket, the following message will
appear: “A-9F: MF instrument disconnected”.
Note:
If non-compatible cables are plugged in, this will lead to the appearance of an “incompatible”
error message.
4-10
12.06 / V1.0
AUTOCON
®
II 400
MODEL 205352 20-1
Содержание AUTOCON II 400
Страница 8: ......
Страница 9: ...AUTOCON II 400 AUTOCON II 400 AUTOCON II 400 20 5352 20 11x 20 5352 20 11x 20 5352 20 11x...
Страница 10: ......
Страница 13: ...III Ger teabbildungen Photographs of the equipment Im genes del equipo...
Страница 14: ...IV W R Q Ger teabbildungen Photographs of the equipment Im genes del equipo I O Y U T E...
Страница 146: ......
Страница 152: ......
Страница 156: ......
Страница 158: ......
Страница 172: ......
Страница 186: ...Figure 4 11 Figure 4 12 Figure 4 13 4 14 12 06 V1 0 AUTOCON II 400 MODEL 205352 20 1...
Страница 192: ......
Страница 198: ......
Страница 200: ......
Страница 202: ......
Страница 203: ......