10.2.2
M-Bus + pulse inputs (67-00-20) (PCB 5550-831)
The M-Bus module is supplied via the M-Bus network and is independent of the meter’s own supply. M-Bus and
the energy meters communicate two-way via opto couplers which gives galvanically separation between M-Bus and
the meter. The module supports primary, secondary and enhanced secondary addressing.
The M-Bus module has 2 extra inputs. See paragraph 7.3 Pulse inputs VA and VB concerning functioning of the
pulse inputs.
In order to function correctly in a MC602, minimum program version H1, released in March 2011, is required.
10.2.3
Radio pulse inputs (67-00-21) (PCB 5550-805)
The radio module is supplied as standard to operate in a licence-free frequency band but can also be supplied to
other frequences requiring licence.
The radio module is prepared to form part of a Kamstrup radio network, where the data are automatically transferred
to system software via the network components RF Router and RF Concentrator.
The radio module has 2 extra inputs. See paragraph 7.3 Pulse inputs VA and VB regarding functioning of the pulse
inputs. The RadioRouter module must be used with mains supply.
Kamstrup recommend that an external antenna is mounted on this module if the meter is fitted with a top module
too. This ensures the best possible radio range.
10.2.4
Prog. data RTC + 4…20 mA pulse inputs (67-00-22) (PCB 5550-925 )
The module has connection possibility for two pressure transmitters on terminals 57, 58 and 59 and can be
adjusted for current reading or pressure ranges of 6, 10 or 16 bar.
The module is prepared for remote reading, data from meter/module being transferred to the system software via
the connected external GSM/GPRS modem on terminals 62, 63 and 64.
The prog. data logger function is configured for the specific meter where it is installed by using the serial number.
If the module is removed and installed in another meter, the configuration is set to default values.
Furthermore, the module has two extra pulse inputs, see section 7.3: Pulse inputs VA and VB as to function. The
module must always be powered by 24 VAC.
Pressure transmitter requirements: 4…20 mA, 2-wire, loop-powered, loop voltage max. 16 VDC
(e.g. type CTL from Baumer A/S)