MULTICAL®
602
114
Kamstrup
A/S
∙
Technical
Description
∙
5512
‐
931_N1_GB_11.2016
13.1.3
Total programming
To do total programming the meter must, via an optical eye, be connected to a PC with the METERTOOL software
running. Break the verification seal and short circuit the TOTAL PROG button on the inside of the calculator top
with a short circuit pen (6699-278).
Note!
This should only be done by an accredited laboratory since breaking of the verification seal voids the legal
verification as well as the factory warranty.
The TOTAL PROG button is activated by short circuiting the two points on the PCB during programming. When the
desired values have been set, METERTOOL terminates programming mode by a reset, and the meter returns to
normal mode ready for use.
Verification seal
TOTAL PROG-button
Figure
13
Figure
14
13.1.4
Time / date (Basic/Advanced Mode)
In this menu the built-in clock in the meter can be read out and adjusted either manually or by setting the meter to
the clock of the PC, on which METERTOOL is running. Having corrected the time, the new time and date can be
written to the meter.
13.1.5
Flow sensor Communication on/off (Advanced Mode)
In this menu the digital communication between flow sensor and meter can be turned on or off – see below.
13.1.6
Modules (Advanced Mode)
This is the module setup menu used for configuration of top and base modules.
13.1.7
Preset VA / VB (Advanced Mode)
If the external pulse inputs (VA and/or VB) are used, the preset values can be entered here.
13.1.8
Print label (Advanced Mode)
If the meter configuration has been read prior to entering this menu, the front label can be printed here.
13.1.9
Verification (Advanced Mode)
This is the verification menu where the calculator can be verified to check if the accuracy is within the given limits.
See verification description in next chapter of this document.
13.1.10
Verification unit settings (Advanced Mode)
In this menu the verification unit settings can be read from the unit altered, and programmed into the verification
unit.
13.1.11
Verification unit calibration (Advanced Mode)
Used for changing between temperature set points during calibration.
13.1.12
Certificate (Advanced Mode)
In this menu previously saved verification results can be printed out in certificate form.