602-C
Application No. 9
2 cooling circuits with joint inlet
Cooling energy #1: E4 = V1(T1-T3)k
T1:Inlet
Cooling energy #2: E5 = V2(T2-T3)k
T2:Inlet
602-C
Application No. 10
Hot tap water energy: E1 = V1 (T1-T2)K
T1:Inlet
T1 is measured with a 2-wire sensor (602-C) or with a
4-wire sensor (602-B/D)
T2 is either measured with a 2-wire sensor (
602
-C) or with a
4-wire sensor (
602
-B/D)
Or
T2 is programmed with a fixed temperature value
Or
T2 is programmed via
the scheduler and hourly datalogger top module,
type 67-0A. The temperature T2 will then follow a schedule
where T2 changes up to 12 times per year.