Full Digital Color Doppler Ultrasonic Diagnostic Instrument (Vet) User’s Manual V1.00
- 72 -
21.1.2 USB Print
1. Insert U disk into the USB port at the back of main unit;
2. When U disk connected to this instrument, the status bar at the bottom right of the screen will pop up
USB icon;
3. Save the images or reports within the system to U disk, unplug the U disk after completed saving;
4. Connect U disk to PC, you can print out the images or reports by the default printer in the PC.
21.2 Transfer BMP files
After successfully set the network, the images or reports can be transferred to PC application software
“Network Printer”.
Operation method:
After successfully set the network, freeze the image, press
key on the keyboard to transfer the
image. When the lower-left corner of the screen shows “Transfer success”, it indicated the image has
been transferred to Network Printer. If appears “Transfer fail” prompt, re-set the network.
21.3 DICOM parameter setting
After the success of the network settings, set the DICOM parameters, ultrasound images can be
transmitted to the DICOM server.
1. Press
key on the keyboard of the instrument to open
【
System Preset
】
menu, move the cursor to
select
『
Net Setting
』
, press
『
Set
』
key to show the “Net Preset” dialog box, as below shown:
Figure: Dicom Server Set
(1) DICOM service IP: for example. Client Dicom server IP is 192.168.0.92, you need to enter IP
192.168.0.92 in the DICOM Server Set.
(2) DICOM service port: is fixed value “104”.
2. After the setting is complete, move the cursor to select “Verify Settings” button marked on the DICOM
Server Set, press
『
Set
』
key, it will pop up success prompt box, press
「
OK
」
in the prompt box, the
DICOM server set is successful, you can transfer DICOM files. If pop the warning box “Service is
timeout, not support Dicom”, please check the IP address of DICOM server and TCP port, as well as
the network cable is properly connected.
21.4 Transfer DICOM files
After successfully set the network, the images or reports can be transferred to DICOM server.
Operation method:
Freeze the current interface, press
key to enter image management work interface. Open the file in
the D disk or U disk, press
〖
Dicom
〗
button, the images or reports saved in the D disk or U disk will be
sent to DICOM server. If prompts “Service is timeout, not support Dicom”, please re-set network and
DICOM server.