Chapter Eleven Cine playback
In frozen status, system automatically enters the manual playback status of cine mode, user toggles the
trackball to view the images frame by frame, can measure, text annotate, body mark and other operations for
the valuable images; press
key to realize the switching between manual playback and automatic
playback.
11.1 Enter/exit cine playback
1. Enter the cine playback
In B, CFM or PDI mode, press freeze key to freeze image, the manual prompt is displayed in the status
prompt region (as shown below), system automatically enters manual loop mode, user can toggles the
trackball to view the playback images frame by frame.
2. Exit the cine playback
In the cine mode, press freeze key again to unfreeze image, return to real-time scanning status and exit
the cine mode.
11.2 Operation of cine playback
11.2.1 Switch automatic and manual playback
In cine mode and in manual playback mode, press
key to realize the switching between manual
playback and automatic playback. Press
key, system enters automatic playback status; press
key again,
system returns to manual playback status.
11.2.2 Storage of image or video
In image frozen status, press
key to realize one-key image storage; in the manual play status,
press
key to realize video storage.
1. Store image
(1) Freeze the image, press
key, after completion of the storage, the prompt appeared at the left
corner of the screen is changed from “Being stored” to “Been stored”; the current displayed
image will be stored in the D disk or U disk.
(2) Freeze the image, press
key to pop up the “Store Option” dialog box; move the cursor to
choose “Image”, press
『
Set
』
key, choose U or D disk, press
『
Set
』
key again, the image will be
saved in the selected disk.
2. Store video
Freeze the image, press
key to pop up the “Store Option” dialog box; move the cursor to choose
“Video”, press
『
Set
』
key, choose U or D disk, press
『
Set
』
key again, the video will be saved in the
selected disk.
Note: In B, CFM or PDI mode, you can store the video.
Attention:
1. Before a new patient examination, press
key to new patient information, clear the current
cine playback, to avoid the confusion of the images for new and old patients.
2. Under the circumstance that the absence of input patient information or the current patient
information deleted, the device will automatically store patient information according to the
system’s date.
3. In PW mode, storage must be stored for more than one frame that can achieve playback, so
storage must reverse long enough time.
4. Operating U disk
and D disk should allow sufficient storage time to prevent the files are missing.
5. When plug U disk, the default storage to U disk; if not plug U disk, the default storage to D disk.
6. Two USB ports at the rear panel of main unit can not be used simultaneously, if use only uses
one of them.