Full Digital Color Doppler Ultrasonic Diagnostic Instrument (Vet) User’s Manual V1.00
- 9 -
5.1 System placement
Please carefully read through and fully understand the safety cautions before moving and placing the
system.
1. Unpack the instrument case and check the goods for its completeness according to the packing list
furnished.
2. Place the instrument on a stable and leveled position.
3. Leave adequate space of 20 centimeters as minimum from rear, left and right side of the instrument.
5.2 Probe brackets installation
The probe bracket should be mounted behind the main unit.
5.3 Ultrasonic probe installation
5.3.1 Probe connection
1. Shutdown the system; open the probe locking knob, and insert the probe connector into the probe
socket at the back of the main unit until the metal spring plates is evenly contacted. Tightly press the
connector.
2. Rotate the probe locking knob 90 degree clockwise and lock it, shown as the figure
.
5.3.2 Probe disconnection
Shutdown the system; rotate the locking knob 90 degree counterclockwise, pull out the probe connector
vertically.
5.4 Connect to power
1. Connect to the power adapter
Insert the output plug of adapter into DC input interface, which is at the back of main unit.
Attention:
Adequate space from rear, left and right side of the machine shall be reserved,
or the machine may malfunction under excessive heat inside the enclosure.
Attention:
1. Turn off the ultrasonic system before disconnecting the probe. Disconnecting the probe
with system power on may damage the system or probe.
2. Before disconnecting the probe, place the probe on the probe bracket so that the probe
may not be damaged or injury person by unexpected fall.
3. Freeze the instrument when instrument is start-up without operation to increase of service
life of probe.
4. Repeat available machine time should be more than 5 minutes to avoid turn on/off power
supply in short time.
Warning:
1. Do not use the probe not provided by our company, otherwise the equipment and the probe
will cause damage, and may cause fire in extreme cases.
2. Check the ultrasonic probe and connecting cable after diagnostic operation. Use of
defective probe may cause electric shock to the patient.
3. Do not strike the probe; using the damaged probe may cause electric shock to the patient.
4. Unauthorized disassembly of the probe shall be prohibited as it may cause electric shock.
Warning:
Before connecting or using the probe, make sure that the probe, connecting
cable and connector are in normal condition (free of cracks or drop). Use of defective
probe may cause electric shock.