Full Digital Color Doppler Ultrasonic Diagnostic Instrument (Vet) User’s Manual V1.00
- 8 -
Chapter Five System Installation and Check
Warning:
1.
All plugs of instruments of this system shall be connected into the power socket with
protectively earth on the wall and the socket must meet the requirement of power rating of
instrument. Use of multiple portable socket-outlets may affect protective earth to make
leakage currents exceed the safety requirements.
2.
Please follow the correct electrical connections method to connect the power supply and
earth, otherwise there will be danger of electric shock. Do not connect the grounding wire to
any gas pipe or water pipe, or it may cause bad grounding and danger of explosion.
3. This equipment is not waterproof, not use this equipment in place where liquid may into the
interior of the equipment. Never pour any liquid on the equipment; otherwise there will be
danger of electric shock or cause equipment damage. If accidentally spill liquid on the
equipment, turn off the power immediately and contact your local representative.
4. Prohibit the live parts of the equipment or other devices (such as various signal input and
output ports, etc.) contact with the patient, if this equipment or other equipment has failure,
the patient will have danger of electric shock.
5.
Additional equipment connected to the medical electrical equipment must comply with the
respective IEC or ISO standards (e.g. IEC60950 for data processing equipment).
Furthermore all configurations shall comply with the requirements for medical electrical
systems (see IEC 60601-1-1 or clause 16 of IEC60601-1 3rd, respectively). Anybody
connecting additional equipment to medical electrical equipment configures a medical
system and is therefore responsible that the system complies with the requirements for
medical electrical systems. Attention is drawn to the fact that local laws take priority over
the above mentioned requirements. When more than one device is connected to use, the
accumulation of leakage current may cause the danger of
security.If in doubt, consult your
local representative or the technical service department.
Warning:
1.
When instrument works abnormally, do stop working, turn off the power and check the
reason, then contacts the KAIXIN Company about it.
2.
Turn off power and pull out of the plug from socket after each ultrasonic diagnostic
operation.
3.
It is forbidden to drag and press the power and probe cables emphatically; regularly
inspect whether there is pull-apart and bareness, if there is the phenomena like this, turn off
power supply immediately, stop using it and change it for new one.
4.
It is forbidden to load and unload the probe or move the instrument in galvanic to avoid
danger of safety.
5.
Pull out of the plug from socket after operation in thunderstorm weather to avoid the
instrument being damaged by lightening.
6.
If the temperature changes greatly in short time will cause vapor recovery inside of
instrument, the case may damage the instrument.
7.
The equipment is switched completely only by disconnecting the power supply from the
wall socket.
Warning:
To avoid the risk of electric shock, this equipment must only be connected to a
supply mains with protective earth.
Warning:
The power adapter, probes, foot switch as described in this section may be
replaced by operator. To replace according to the method of this section, does not result in
an unacceptable risk.