Full Digital Color Doppler Ultrasonic Diagnostic Instrument (Vet) User’s Manual V1.00
- 10 -
2. Connect to the main power supply
Insert the power plug (jack) furnished with the machine into power input socket of the power adapter,
the other end to the mains socket-outlet. The instrument uses three-core power supply. It connects with
the protective earth line when power plug inserts into its socket.
5.5 Install foot switch
Shutdown the system; insert the plug of foot-switch into the “
” socket at the back of the main unit.
5.6 Video recorder installation
1. Connect the equipotential terminal ( ) of the video recorder to the earthing;
2. Connect one end of the video connection line to the video recorder and the other end to the video output
interface (
) on the real panel of main unit;
3. Insert one end of power plug (jack) of the video recorder to the power input socket of the video recorder,
the other end to the power supply socket.
5.7 Ultrasonic probe check before and after operation
Check if there are any exceptionally on the surface of the probe or cable jacket, such as peeling, cracks,
bulge, or if the acoustic lens is reliable, disinfected or cleaned.
5.8 Main unit check before and after operation
5.8.1 Inspection before start-up
Check the following items before starting the machine:
1. The temperature, humidity and atmospheric pressure shall meet the requirements of operation
condition.
2. No condensation occurs.
3. No distortion, damage or contamination on system and peripheral. Clean the parts as specified in
relevant sections, if the contaminant is present.
4. Check the control panel, LED screen and enclosure to ensure they are in good working condition and
free of abnormity (such as cracks and loosened screws).
5. No damage on cable (power cable, etc.), and hard up on its connection.
6. Check probe and its connections to ensure they are free of abnormity (such as scuffing, drop-off or
contamination). If the contaminant is present, clean, disinfect the contaminated objects as specified in
relevant sections.
7. See to it that there are no foreign objects on the control panel (Remove the object if any).
8. No barriers around the intake of equipment.
9. See to it that probe has been cleaned, disinfected; else dispose it as specified in relevant sections.
10. Check all the ports of the machine for possible damage or blockage.
11. Clean the field and environment.
Warning:
1. Adapter has no switch. APPLIANCE COUPLER or MAINS PLUG is used as the isolation
means from the SUPPLY MAINS. Not to position the EQUIPMENT so that it is difficult to
operate the disconnection device.
2. The device should be used only with power adapter provided by Kaixin Company.
3. To avoid damaging power adapter or harming people by unexpected fallen, make sure the
power adapter is placed on the leveled plane.
4. The operator must not touch signal input/signal output and patient simultaneously.
Attention:
The waterproof grade of foot switch is IPX1.