Full Digital Color Doppler Ultrasonic Diagnostic Instrument (Vet) User’s Manual V1.00
- 15 -
1.
Move the cursor to the “Waiting Time” input box, press
『
Set
』
key to confirm the cursor position;
2.
Input the screensaver time, number (0-99), in minute.
Note: Go beyond the system setting energy-saving time without pressing any key, the system will
automatically enter the energy saving status. Press any key, system will return to normal
operation status.
Key Sound setting
Move the cursor to the “Key Sound” select box, press
『
Set
』
key again and again., choose “Close” and
“Open” which showed alternately in the select box.
After presetting the system data “Hospital”, “Language Set” “TV Mode”, “Font Bright”, “Time Set”,
“Date Set”, “Date Format”, “Waiting Time” and “Key Sound”, move the cursor to
「
OK
」
, press
『
Set
』
key,
the system will preserve the data of preset and exit the dialog box; If choose
「
Cancel
」
, press
『
Set
』
key, the
system will give up the data of preset and exit the dialog box. You also can press
key to give up and exit
the preset.
Note: If you modify the core data “Language Set”, “TV Mode” or “Font Bright” in the basic
preset, please reboot and save the settings.
7.3 Probe Information
Move the cursor to the
『
Probe Info
』
, press
『
Set
』
key, the screen will pop up the “Probe Information” dialog
box. The dialog box displays all the probes supported within the system and the relative information about each
probe. Move the cursor to
「
OK
」
, press
『
Set
』
key to exit the probe information dialog box.
7.4 Parameter Display
Move the cursor to the
『
Param Display
』
, press
『
Set
』
key, the screen will pop up the “System default
parameter display” dialog box. The details see
Chapter Nine “9.2 User Settings”
.
7.5 Formula preset
Move the cursor to the
『
Formula
』
, press
『
Set
』
key, the screen will pop up the “OB Formulas Preset” dialog
box.
Move the cursor to the gestational week position of the bovine, canine, equine, feline, goat, llama, ovine
and swine, press “Reset”; then choose goat species or obstetric table for canine, move the cursor to
「
OK
」
after finishing the preset and press
『
Set
』
key, the system will preserve those data of preset and exit the dialog
box; If you choose
「
Cancel
」
and press
『
Set
』
key, the system will give up those data of preset and exit the
dialog box. You also can press
key to give up and exit the preset.
7.6 Net Setting
1. Move the cursor to the
『
Net Setting
』
, press
『
Set
』
key, the screen will pop up the “Net Preset” dialog box.
The details of Net Setting see
Chapter Twenty-one “Print and Transfer files”
.
2. Choose
「
Cancel
」
press
『
Set
』
key or directly press
key to exit the net preset.
7.7 Comment preset
1. Move the cursor to the
『
Comment
』
, press
『
Set
』
key, the screen will pop up the “Comment Preset” dialog
box.
Comment preset is divided into seven kinds: lesions, general, abdomen, obstetric, cardiac, urology and
superficial. Each comment is set common noun, user can add or modify the comments according to the actual
need.
①
New: Add the contents of the input box to the annotation list box; if repetition, not add;
②
Edit: The contents in the input box are updated to the annotation list box;
Warning:
After
software upgraded, start the machine and must first click “Reset” in the
“OB Formulas Preset” to restore the default values; otherwise obstetric table may occur
calculation error.