9 Configuring the transmitter
9
Configuring the transmitter
9.1
Startup level (In)
In this level, the measured values are measured and transmitted quickly. The measured values are taken
every 15 s and transmitted every second, even with a USB connector inserted. If the configured mea-
suring interval is less than 15 s, the shorter interval is used. At the shortest measuring interval, the CO
2
sensor measures every 30 s.
NOTE!
If a USB connector is inserted, all measurements are carried out except the CO
2
measurement. The CO
2
measured values are "frozen" in the inserted USB connector. If no measured value is available, "- - - -"
is displayed on the LCD display alongside the measured value number and measurand.
NOTE!
It is
not
possible to take a CO
2
measurement while a USB connector is inserted.
Fig. 9-1 Sample appearance for the startup level
1. Hold the
key for > 2 s to change from the normal display to the startup level.
Startup level is activated.
2. Select the measured values of the measuring probe using the
or
keys.
The measured values are taken every 15 s and transmitted every 1 s, even with a USB connector
inserted. If the configured measuring interval is less than 15 s, the shorter interval is used. At the
shortest measuring interval, the CO
2
sensor measures every 30 s.
3. To return to the normal display, hold the
key for > 2 s or use the 2 min key timeout.
from NA
> 2 s (In)
to NA
> 2 s
to NA
> 2 s
to NA
> 2 s
RH
RH