37
7 Introduction to operation
7.3
Startup level (In) (measuring and transmitting measured values quickly)
The startup level comes in useful if, for example, you wish to determine the optimum antenna position
when mounting the device, as this level allows measured values to be "measured and sent quickly".
By changing to the startup level (In), "InbE" and the measured value of 1 are displayed alternately. In
addition, the measured value number and the corresponding measurand may also appear.
The display changes after 2 minutes if none of the operating keys are pressed. Once this time has
passed, it will automatically return to the normal display (NA).
The measured values are taken every 15 s and transmitted every 1 s, even with a USB connector insert-
ed. If the configured measuring interval is less than 15 s, the shorter interval is used. At the shortest mea-
suring interval, the CO
2
sensor measures every 30 s.
NOTE!
If a USB connector is inserted, all measurements are carried out except the CO
2
measurement. The CO
2
measured values are "frozen" in the inserted USB connector. If no measured value is available, "- - - -"
is displayed on the LCD display alongside the measured value number and measurand.
NOTE!
It is
not
possible to take a CO
2
measurement while a USB connector is inserted.
Keys and key combinations
RH
RH
Keys
Function
or
Selection of measured values 1 to 4
> 2 s
Return to the normal display (NA)