17
3 Preparing the measuring probe
3
Preparing the measuring probe
CAUTION!
Explosion of the batteries or fire hazard.
Supply is either via batteries or external voltage supply.
Simultaneous supply via batteries and external voltage supply not possible.
3.1
Receiver software version
NOTE!
The measuring probe requires a receiver with a software version (firmware) of 05.01 or higher.
3.2
Supply with batteries
3.2.1
Inserting/changing the battery
CAUTION!
Make sure that pollutants, moisture, and steam cannot enter the measuring probe.
The measuring probe could be destroyed.
When inserting/changing the batteries make sure that the measuring probe is not exposed to pollut-
ants, moisture, and steam. In addition, it must be ensured that the seal is correctly fitted between the
battery compartment and measuring probe.
CAUTION!
The measuring probe does not function if the poles of the batteries are incorrectly connected.
The transmitter's batteries and electronics could become damaged.
Make sure that the poles of the batteries are correctly connected.
CAUTION!
Incorrect batteries put safety at risk.
The measuring probe may be destroyed if incorrect batteries are used.
It is preferable that you only use the lithium batteries recommended as accessories.
CAUTION!
Electrostatic discharge
The measuring probe is fitted with electronic components that may be destroyed by electrostatic dis-
charge.
When opening the measuring probe to change the battery or connect the external voltage supply, it
must be ensured that the personnel are sufficiently electrostatically discharged.