41
8 Operating the measuring probe
8.3
Changing to different levels
Fig. 8-1 Changing to different levels
Hold the
key for > 2 s to change to the startup level (In). Here the measured values are taken every
15 s and transmitted every 1 s, even with a USB connector inserted. If the configured measuring interval
is less than 15 s, the shorter interval is used. At the shortest measuring interval, the CO
2
sensor mea-
sures every 30 s.
Hold the
key for > 2 s to change to the parameter level (PA). All functions of the measuring probe,
the general parameters of the measuring probe, and the specific parameters for the measured values
are determined here.
Key timeout
If no key is pressed in the startup level for a period of 2 min and or 40 s in the parameter level, the mea-
suring probe automatically returns to the normal display.
Code request
The transmitter features one code request for each change to the parameter level. However, this code
request is not active when delivered. In the parameter level, a code (minimum 1 digit, maximum 4 digits)
can be assigned.
chapter 8.4 "Code request", Page 42
> 2 s (PA) (code inquiry)
> 2 s (In)
> 2 s or timeout 2 min
> 2 s or timeout 40 s
Standard display (NA)
RH
Parameter level (PA)
RH
RH
Start-up level (In)