49
9 Configuring the transmitter
Measured values 1 to 4
The following table shows the parameters to be selected for measured value 1. The same setting options
can also be used for the measured values 2 to 4.
Parameter
Display
Value range/selection
Selecting the measurand
If activation for the measurands
(value range/selection 6 to 12) is
not available, then these are not
available for selection and are also
not displayed.
dAtA
0 = No measured value
1
= Measured value for automati-
cally recognized sensor type (Plug
and Play)
2 = Pt1000 temperature [°C]
3 = CO
2
[ppm CO
2
]
4 = Relative humidity [%RH]
5 = Temperature [°C]
6 = Absolute humidity [g/m
3
]
7 = Dew point temperature [°C]
8 = Mixing ratio [g/kg]
9 = Steam pressure [mbar]
10 = Steam pressure [hPa]
11 = Wet-bulb temperature [°C]
12 = Enthalphy [kJ/kg]
Measuring interval factor
(Measuring interval =
Transmission interval × Measuring
interval factor)
This is not displayed if the mea-
surand has been configured to "no
measured value".
Restrictions:
Sensor for temperature (Pt1000)
Min. measuring interval 1 s
Default measuring interval 15 s
Sensor for humidity
Min. measuring interval 5 s
Default measuring interval 90 s
Sensor for CO
2
Min. measuring interval 30 s
Default measuring interval 1200 s
If Plug and Play is active, this pa-
rameter has no function.
FAct
1 to 30
to 3600
Selection of the sensor channel
This is not displayed if the mea-
surand has been configured to "no
measured value".
SEnS
1 to 2
Offset (Process value correction)
This is not displayed if the mea-
surand has been configured to "no
measured value".
oFFS
-99.9 to
0.0
to +99.9
Display measured value
Do not display measured value
This is not displayed if the mea-
surand has been configured to "no
measured value".
SHo
on
oFF