3 Preparing the measuring probe
20
3.2.3
Battery operating life
The expected battery operating life can be determined with the calculation program in the setup program
(accessories) or with a 30-day test version (free) for lithium and alkaline batteries.
The battery operating life can only be calculated with fixed configured measurands in measured values
1 to 4. The operating life cannot be calculated with automatic sensor detection.
The low bat alarm time is approximately 2 % of the calculated battery life.
The measuring probe's battery operating life is calculated for room temperature and depends on the bat-
tery type, sensor type, transmission interval, measuring interval, and ambient temperature.
Operating the probe at the limit of the battery's admissible ambient temperature or in environments with
frequent temperature fluctuations can reduce the batteries' actual operating life by up to 50 %.
Leaving the backlight switched on permanently reduces the battery operating life to a few days.
3.2.4
Disposal of batteries
Please dispose of all batteries as specified by the legislator according to the German Closed Substance
Cycle and Waste Management Act or local regulations.
The contacts of batteries that are not fully discharged electrically need to be insulated. Disposal of bat-
teries together with domestic waste is expressly prohibited. Batteries can be taken to municipal waste
collection points or local retailers at no extra charge.