Capter 3
Function Setting and Basic Operation
Curve Tracer CS-3000 Series Instruction Manual
3-23
3.6.3 Setting OFFSET knob
3.6.3.1
OFFSET knob and display menu
Rotating OFFSET knob
③
allows the offset voltage or current value of 0 step to be set.
Setting range and functions
The range from
−
10 times to
+
10 times of voltage or current per step of the step generator can
be set.
The offset value changes together with STEP AMPLITUDE 1-2-5 step.
If STEP AMPLITUDE
=
100mV/div and the offset
=
100mV, changing STEP AMPLITUDE to
200mV/div or 500mV/div makes the offset values change to 200mV or 500mV too.
Pressing OFFSET knob allows the setting value to be cleared to 0.
The resolution is 1
%
of the voltage or current per step.
The offset setting is not affected by STEP 1/10 key.
The menu above shows the setting values.
0 step
Offset