Capter 3
Function Setting and Basic Operation
3-36
Curve Tracer CS-3000 Series Instruction Manual
SWEEP STEPS
- The setting range of SWEEP STEPS is from 20 to 1000. However, the actual number of measurement
points uses the smaller number between SWEEP STEPS setting value and 10 times of VARIABLE setting
value.
- This is because division by the minimum resolution (0.1
%
) or more of VARIABLE cannot be made.
However, generally the range from VARIABLE setting value up to 0
%
is divided by the number of SWEEP
STEPS when being measured.
Starting SWEEP measurement
- When the collector supply SWEEP is available, pressing SWEEP key allows the collector supply SWEEP
measurement to start.
During the collector supply SWEEP, the state is displayed as shown below:
SWEEP
During collector supply SWEEP
3.2
%
Current VARIABLE value
It ends the measurement when you acquire data within the range of VARIABLE corresponding to
the setting of SWEEP TYPE. "Sweep completed" is temporarily displayed on the left of the screen,
and it enters the state of STOP.
Stopping and restarting SWEEP measurement
Push REPEAT key or STOP key to stop SWEEP.
When SWEEP is completed, VARIABLE setting value returns to the value when SWEEP starts.
Push SWEEP key after setting an appropriate VARIABLE value again when restarting.
For example, if SWEEP STEPS
=
500 and VARIABLE setting
=
45.2
%
, the number of SWEEP
measurement points should be 452.
Содержание CS-3000 Series
Страница 1: ...Instruction Manual Curve Tracer CS 3000 series CS 3100 CS 3200 CS 3300 ...
Страница 2: ...Ⓒ 2010 IWATSU ELECTRIC CO LTD All rights reserved ...
Страница 16: ......
Страница 170: ......
Страница 210: ...Chapter 4 Characteristics Measurement 4 40 Curve Tracer CS 3000 Series Instruction Manual Memo ...
Страница 214: ...Chapter 5 Daily Maintenance and Calibration 5 4 Curve Tracer CS 3000 Series Instruction Manual Memo ...
Страница 221: ...Chapter 6 Specifications Curve Tracer CS 3000 Series Instruction Manual 6 7 6 3 3 Test fixture CS 302 Unit mm ...
Страница 223: ...Chapter 6 Specifications Curve Tracer CS 3000 Series Instruction Manual 6 9 6 3 5 Test fixture CS 301 190 Unit mm ...