Capter 3
Function Setting and Basic Operation
3-48
Curve Tracer CS-3000 Series Instruction Manual
3.10.2 INVERT key
Each press of INVERT key
②
changes inverse/non-inverse display of a waveform.
In non-inverse display (normal state), LED beside INVERT key goes off and it goes on in inverse
display.
VIEW MODE and GND level display
∗
For details of VIEW MODE, see Section 3.10.3.
Origin for TRACE and WAVE
TRACE
All things in the waveform display area are inverted horizontally and vertically; i.e., 180
°
rotation
using the center of the screen as standard.
The origin (GND level), waveform, and cursor are inverted. REF waveform is not inverted.
WAVE
Waveforms in WAVE 1st and 2nd are inverted vertically.
When not being inverted, the position of the origin depends on polarity settings of the collector
supply and the step generator. (When the polarity is positive, the origin is at lower or left
position. When it is negative, the origin is at upper or right position.)
TRACE display
WAVE display
INVERT off
Origin
INVERT off
INVERT on
INVERT on
GND level mark V
GND level mark V
GND level mark V
GND level mark V/H
GND level mark H
Origin
Inverted display
Inverted display
Содержание CS-3000 Series
Страница 1: ...Instruction Manual Curve Tracer CS 3000 series CS 3100 CS 3200 CS 3300 ...
Страница 2: ...Ⓒ 2010 IWATSU ELECTRIC CO LTD All rights reserved ...
Страница 16: ......
Страница 170: ......
Страница 210: ...Chapter 4 Characteristics Measurement 4 40 Curve Tracer CS 3000 Series Instruction Manual Memo ...
Страница 214: ...Chapter 5 Daily Maintenance and Calibration 5 4 Curve Tracer CS 3000 Series Instruction Manual Memo ...
Страница 221: ...Chapter 6 Specifications Curve Tracer CS 3000 Series Instruction Manual 6 7 6 3 3 Test fixture CS 302 Unit mm ...
Страница 223: ...Chapter 6 Specifications Curve Tracer CS 3000 Series Instruction Manual 6 9 6 3 5 Test fixture CS 301 190 Unit mm ...