Capter 3
Function Setting and Basic Operation
Curve Tracer CS-3000 Series Instruction Manual
3-59
3.10.6 POSITION key
Pressing POSITION key
⑥
allows "POSITION" menu to appear.
The origin of the characteristic curve can be set on an arbitrary position.
Rotating FUNCTION knob allows each position to be set.
Pushing FUNCTION knob allows the position to be set to 0.00div.
F1
Sets the vertical position.
F2
Sets the horizontal position.
F3 to F6
(Not used)
POSITION movable area and resolution (Zoom: OFF)
- Setting resolution: 0.02 div
- The movable areas are as follows, depending on display range.
- For Invert setting, the move direction and rotation direction of FUNCTION knob are inverse.
COLLECTOR
Collector supply MODEs
AC
Positive polarity other than AC Negative polarity other than AC
-5.00 to +5.00 div
Zero position
=
center of screen
0.00 to 10.0 div
Zero position
=
left edge or
lower edge of screen
−
10.0 to 0.00 div
Zero position
=
right edge or
upper edge of screen
EMITTER
Collector supply MODEs
Positive
Negative
0.00 to 10.0 div
Zero position
=
left edge of screen
−
10.0 to 0.00 div
Zero position
=
right edge of screen
BASE-EMITTER
STEP GENERATOR MODE
Positive
Negative
0.00 to 10.0 div
Zero position
=
left edge of screen
−
10.0 to 0.00 div
Zero position
=
right edge of screen
STEP
−
5.00 to
+
5.00 div
This setting is possible but the display is fixed at the specified position; cannot be
moved.
F1
F2
Содержание CS-3000 Series
Страница 1: ...Instruction Manual Curve Tracer CS 3000 series CS 3100 CS 3200 CS 3300 ...
Страница 2: ...Ⓒ 2010 IWATSU ELECTRIC CO LTD All rights reserved ...
Страница 16: ......
Страница 170: ......
Страница 210: ...Chapter 4 Characteristics Measurement 4 40 Curve Tracer CS 3000 Series Instruction Manual Memo ...
Страница 214: ...Chapter 5 Daily Maintenance and Calibration 5 4 Curve Tracer CS 3000 Series Instruction Manual Memo ...
Страница 221: ...Chapter 6 Specifications Curve Tracer CS 3000 Series Instruction Manual 6 7 6 3 3 Test fixture CS 302 Unit mm ...
Страница 223: ...Chapter 6 Specifications Curve Tracer CS 3000 Series Instruction Manual 6 9 6 3 5 Test fixture CS 301 190 Unit mm ...