Chapter 4 Characteristics Measurement
Curve Tracer CS-3000 Series Instruction Manual
4-37
(13) Press CONFIGURATION key.
In HIGH CURRENT, the setting is fixed to Collector/StepGen/Common
Checking pulse waveform in WAVE mode
(14) Press VIEW/PULSE key.
(15) Press F1 key to set VIEW MODE to WAVE.
WAVE 1st
(16) Press F2 key to set to Vce.
(17) Rotate FUNCTION knob to set Vce range. Make proper setting in accordance with device
measurement value.
WAVE 2nd
(18) Press FUNCTION knob to set to Ic.
(19) Rotate FUNCTION knob to set Ic range. Make proper setting in accordance with device
measurement value.
(20)Set PULSE WIDTH to 400μs. Make proper setting depending on the device.
(21) Set MEASUREMENT POINT to 300μs. Make proper setting depending on the device.
(22) Move the cursor to the measurement point where waveforms of WAVE 1st and WAVE 2nd
are stable.
Starting measurement
(23) Set HIGH CURRENT breaker to ENABLED.
(24) Gradually rotate VARIABLE and apply the current to the device.
Содержание CS-3000 Series
Страница 1: ...Instruction Manual Curve Tracer CS 3000 series CS 3100 CS 3200 CS 3300 ...
Страница 2: ...Ⓒ 2010 IWATSU ELECTRIC CO LTD All rights reserved ...
Страница 16: ......
Страница 170: ......
Страница 210: ...Chapter 4 Characteristics Measurement 4 40 Curve Tracer CS 3000 Series Instruction Manual Memo ...
Страница 214: ...Chapter 5 Daily Maintenance and Calibration 5 4 Curve Tracer CS 3000 Series Instruction Manual Memo ...
Страница 221: ...Chapter 6 Specifications Curve Tracer CS 3000 Series Instruction Manual 6 7 6 3 3 Test fixture CS 302 Unit mm ...
Страница 223: ...Chapter 6 Specifications Curve Tracer CS 3000 Series Instruction Manual 6 9 6 3 5 Test fixture CS 301 190 Unit mm ...