Capter 3
Function Setting and Basic Operation
3-78
Curve Tracer CS-3000 Series Instruction Manual
⑦
Press the function key so that Software version for update is selected (e.g., press F2 key on the
diagram on the previous page).
⑧
Press F6 key to select SELECT (on the left in the diagram below).
The software of this instrument to be updated is displayed in File List (on the right in the
diagram below).
⑨
Press F6 key to select UPDATE. CONFIRMATION screen appears as shown below.
Select [OK] to execute update.
Select [CANCEL] to display SYSTEM TOOLS submenu without execution of update.
Note 1: When updating the software in the instrument, never set the power supply to standby or remove
the USB memory.
Note 2: Updating the software in the instrument causes all the internal settings to be initialized.
Necessary data of SETUP (setting conditions), TRACE (waveform data), and REF (reference
waveform data) should be saved in the USB memory using SAVE function (see Sections 3.11.3
and 3.11.4) in SAVE/RECALL menu. For initialized items, see the default setting items in "Table
3.11.1: Recall function" and "Table 3.11.2 Not recall function" of Section 3.11.8.
F1
F2
F3
F4
F5
F6
Содержание CS-3000 Series
Страница 1: ...Instruction Manual Curve Tracer CS 3000 series CS 3100 CS 3200 CS 3300 ...
Страница 2: ...Ⓒ 2010 IWATSU ELECTRIC CO LTD All rights reserved ...
Страница 16: ......
Страница 170: ......
Страница 210: ...Chapter 4 Characteristics Measurement 4 40 Curve Tracer CS 3000 Series Instruction Manual Memo ...
Страница 214: ...Chapter 5 Daily Maintenance and Calibration 5 4 Curve Tracer CS 3000 Series Instruction Manual Memo ...
Страница 221: ...Chapter 6 Specifications Curve Tracer CS 3000 Series Instruction Manual 6 7 6 3 3 Test fixture CS 302 Unit mm ...
Страница 223: ...Chapter 6 Specifications Curve Tracer CS 3000 Series Instruction Manual 6 9 6 3 5 Test fixture CS 301 190 Unit mm ...