Capter 3
Function Setting and Basic Operation
Curve Tracer CS-3000 Series Instruction Manual
3-113
STEP GENERATOR SETUP submenu
Pressing F2 key allows STEP GENERATOR SETUP submenu to appear.
Key/knob
Setting items
Function description
F1
Polarity
F1 is used to select the polarity (
+
,
−
) for MINIMUM, MAXIMUM, START
of the step generator. The setting should be consistent with the control
polarity of the step generator.
F2
FUNCTION
MINIMUM
Sets the minimum value (absolute value) of SWEEP range of STEP
GENERATOR.
The setting range and setting resolution depend on STEP range.
- Setting range: 10 times of STEP range
- Setting resolution: 1/100 of STEP range
F3
FUNCTION
MAXIMUM
Sets the maximum value (absolute value) of SWEEP range of STEP
GENERATOR.
The setting range and setting resolution depend on STEP range.
- Setting range: 10 times of STEP range
- Setting resolution: 1/100 of STEP range
F4
FUNCTION
RESOLUTION
Sets the SWEEP resolution of STEP GENERATOR. If this option is
used for operation, the resolution in this setting is used for SWEEP.
- Setting range: STEP range
×
(0.01 to 1)
F5
FUNCTION
START OFFSET Sets START value of SWEEP of STEP GENERATOR.
- Setting range: MINIMUM value to MAXIMUM value
- Setting resolution: 1/100 of STEP range
F6
-
(Not used)
F1
F2
F3
F4
F5
F6
Содержание CS-3000 Series
Страница 1: ...Instruction Manual Curve Tracer CS 3000 series CS 3100 CS 3200 CS 3300 ...
Страница 2: ...Ⓒ 2010 IWATSU ELECTRIC CO LTD All rights reserved ...
Страница 16: ......
Страница 170: ......
Страница 210: ...Chapter 4 Characteristics Measurement 4 40 Curve Tracer CS 3000 Series Instruction Manual Memo ...
Страница 214: ...Chapter 5 Daily Maintenance and Calibration 5 4 Curve Tracer CS 3000 Series Instruction Manual Memo ...
Страница 221: ...Chapter 6 Specifications Curve Tracer CS 3000 Series Instruction Manual 6 7 6 3 3 Test fixture CS 302 Unit mm ...
Страница 223: ...Chapter 6 Specifications Curve Tracer CS 3000 Series Instruction Manual 6 9 6 3 5 Test fixture CS 301 190 Unit mm ...