Chapter 4 Characteristics Measurement
Curve Tracer CS-3000 Series Instruction Manual
4-7
4.1.1.3
Measuring reverse characteristics (breakdown) of diode
, ,
The diagram on the right in this frame shows
the measurement result of the reverse
characteristics based on the diode
performance table below.
In addition, extracts from the performance
table of device characteristics which should be
investigated in advance and the connection
diagram indicating how to connect a
measured device to the attached test adapter
are shown.
Extracts from diode performance table
Absolute max. rating
Peak reverse voltage V
RM
70 V
Peak forward current I
FM
450 mA
Permissible loss
I
O
250mW
Electric characteristics
Item
Measurement Measurement
conditions
value/Max.
Reverse V
R
=60 V
0.1μA
current I
R
Caution!
When starting or finishing measurement, operate CS-3000 as shown below:
When starting measurement, set HIGH VOLTAGE breaker to DISABLED and open the test fixture.
After connecting a measured device, set HIGH VOLTAGE breaker to ENABLED.
After finishing measurement, rotate VARIABLE counterclockwise to set the value to 0% (to discharge
applied voltage).
Set HIGH VOLTAGE breaker to DISABLED.
Connection diagram
Connect the anode of the diode to the collector terminal of
the test adapter and the cathode to the emitter terminal.
For connection between the test fixture and CS-3000, see
Section 3.3.2.
Diode reverse characteristics
Test adapter
Patch panel
Anode
(collector)
Cathode
(emitter)
Содержание CS-3000 Series
Страница 1: ...Instruction Manual Curve Tracer CS 3000 series CS 3100 CS 3200 CS 3300 ...
Страница 2: ...Ⓒ 2010 IWATSU ELECTRIC CO LTD All rights reserved ...
Страница 16: ......
Страница 170: ......
Страница 210: ...Chapter 4 Characteristics Measurement 4 40 Curve Tracer CS 3000 Series Instruction Manual Memo ...
Страница 214: ...Chapter 5 Daily Maintenance and Calibration 5 4 Curve Tracer CS 3000 Series Instruction Manual Memo ...
Страница 221: ...Chapter 6 Specifications Curve Tracer CS 3000 Series Instruction Manual 6 7 6 3 3 Test fixture CS 302 Unit mm ...
Страница 223: ...Chapter 6 Specifications Curve Tracer CS 3000 Series Instruction Manual 6 9 6 3 5 Test fixture CS 301 190 Unit mm ...