Capter 3
Function Setting and Basic Operation
3-10
Curve Tracer CS-3000 Series Instruction Manual
3.3.5 Connecting the measured device with the test adapter CS-501A (Option)
It is not possible to measure at the same time though two kinds of devices to be measured can be
installed in this test adaptor CS-501A. Measure by installing one device to be measured and one
insulation tab.
[Connection procedure]
(1) Open the cover of the test fixture.
If the cover is opened, this instrument is changed to INTERLOCK state and its output is shut
off.
(2) Connect the test adapter CS-501A with the measured device.
1) Select socket A or B depending on the type of the measured device as shown below:
2) Insert the terminals of the measured device into the connection grooves of the socket. Insert
the insulation tab in the socket that becomes empty. (When buying it, the insulation tab is
inserted in socket A or B. )
- For transistor: Connect the collector, base, and emitter to the collector, base, and emitter of
the test adapter.
- For FET:
Connect the drain, gate, and source to the collector, base, and emitter of the
test adapter.
- For diode:
Connect the anode and cathode to the collector and emitter of the test
adapter.
(3) Mount the test adapter on the test fixture.
When mounting the test adapter, push it into the text fixture vertically while keeping its rear
surface parallel to the mounting part of the patch panel.
(4) If the cover of the test fixture is closed after connecting the test adapter, INTERLOCK state is
released and measurement can be carried out.
Caution!
Base resistance and insulation tab
The test adapter CS-501A is equipped with
the short terminal for the base resistance as
an accessory. It can be screwed at both ends
of it.
The resistance with a proper constant is
installed to the base resistance mainly for
protection from oscillation. If an influence by
oscillation is made, install the resistor
element. If not, surely install the short
terminal attached.
In test adaptor CS-501A, insert an accessory
insulation tab in socket A or B not to connect
the measurement element for an accurate
measurement while measuring.
Socket A
Socket B
Содержание CS-3000 Series
Страница 1: ...Instruction Manual Curve Tracer CS 3000 series CS 3100 CS 3200 CS 3300 ...
Страница 2: ...Ⓒ 2010 IWATSU ELECTRIC CO LTD All rights reserved ...
Страница 16: ......
Страница 170: ......
Страница 210: ...Chapter 4 Characteristics Measurement 4 40 Curve Tracer CS 3000 Series Instruction Manual Memo ...
Страница 214: ...Chapter 5 Daily Maintenance and Calibration 5 4 Curve Tracer CS 3000 Series Instruction Manual Memo ...
Страница 221: ...Chapter 6 Specifications Curve Tracer CS 3000 Series Instruction Manual 6 7 6 3 3 Test fixture CS 302 Unit mm ...
Страница 223: ...Chapter 6 Specifications Curve Tracer CS 3000 Series Instruction Manual 6 9 6 3 5 Test fixture CS 301 190 Unit mm ...