Capter 3
Function Setting and Basic Operation
Curve Tracer CS-3000 Series Instruction Manual
3-95
Executing "RECALL DEFAULT"
[Procedure 2]
(1) Press SAVE/RECALL key. SAVE/RECALL/DELETE menu appears.
(2) Press F2 key on the right of the submenu "RECALL."
(3) Press F4 key on the right of the submenu "RECALL DEFAULT."
SETUP (setting conditions) is set to defaults.
For detailed DEFAULT settings, see "Table 3.11.1 Recall functions" in Section 3.11.8 "Default
settings."
[Procedure 3] Recalling TRACE data from USB memory to screen
(1) Press SAVE/RECALL key. SAVE/RECALL/DELETE menu appears.
(2) Press F2 key on the right of the submenu "RECALL."
(3) Press F1 key on the right of the submenu "SETUP/TRACE/REF" and select "TRACE."
(4) Press F3 key on the right of the submenu "FILE LIST."
SETUP information is displayed in the submenu "USB File list" and the waveform display area.
(5) Rotate FUNCTION knob or press F1(
↑
)/F2(
↓
) key to select the file.
(6) Press F6 key on the right of the submenu "SELECT."
The selected file name is displayed in the submenu "FILE LIST."
(7) Press F6 key on the right of the submenu "RECALL" and execute "RECALL."
(8) "Recall from ABCD0001.WFM" is displayed in the message area on the upper part of the
screen and TRACE waveform is displayed in the waveform display area.
At the same time, the menu screen disappears.
(3) (4)
(5) (6)
(7) (8)
Содержание CS-3000 Series
Страница 1: ...Instruction Manual Curve Tracer CS 3000 series CS 3100 CS 3200 CS 3300 ...
Страница 2: ...Ⓒ 2010 IWATSU ELECTRIC CO LTD All rights reserved ...
Страница 16: ......
Страница 170: ......
Страница 210: ...Chapter 4 Characteristics Measurement 4 40 Curve Tracer CS 3000 Series Instruction Manual Memo ...
Страница 214: ...Chapter 5 Daily Maintenance and Calibration 5 4 Curve Tracer CS 3000 Series Instruction Manual Memo ...
Страница 221: ...Chapter 6 Specifications Curve Tracer CS 3000 Series Instruction Manual 6 7 6 3 3 Test fixture CS 302 Unit mm ...
Страница 223: ...Chapter 6 Specifications Curve Tracer CS 3000 Series Instruction Manual 6 9 6 3 5 Test fixture CS 301 190 Unit mm ...