Chapter 4 Characteristics Measurement
Curve Tracer CS-3000 Series Instruction Manual
4-35
Connection diagram
The diagram below shows how to directly connect a measured device to the test fixture.
Connect the gate, emitter, and collector terminals of IGBT to the base, emitter, and collector
terminals of the test fixture using the attached wires.
Connection procedure
(1) Remove the test adapter mounting part from the test fixture.
(2) Set the measured device IGBT in it.
(3) Connect the collector of the measured device IGBT to SENCE terminal, the emitter to SENCE
terminal, and the gate terminal to 300Vpk, using the attached wires as shown below:
Caution!
If SENSE of HIGH CURRENT is not connected, CS-3000 displays and measures nothing in HIGH
CURRENT mode.
To improve accuracy, connect the emitter sense.
Test Lead for
HIGH
CURRENT
Remove the patch panel.
Collector Emitter
Gate
Measured device IGBT