Capter 2
Name and Function of Each Part (Outline)
2-26
Curve Tracer CS-3000 Series Instruction Manual
Caution!
-
Do not apply a voltage exceeding 3 kV or a current exceeding 40 A to each terminal of the
test adapter CS-508.
If 3 kV or 40 A is exceeded, test adapter or a measured device might be damaged.
No.
Name
Description
Ref.
①
Socket for element
Fix, and connect the element of the SMD CHIP type like
figure 2.25.
Section
3.3.9
①
-A Element installation part
Put the SMD CHIP type element.
-
①
-B Knob with spring
Pull the part of the knob right, and place the element by the
power of the spring.
-
①
-C Positional fixed knob
When a knob alone with the spring doesn't wake up, an
upward screw (white) is loosened, the knob is pulled left,
and the position is adjusted, and tighten and fix the screw
(white).
-
②
③
COLLECTOR/DRAIN/ANODE
SENSE
Connect it with the terminal that corresponds to
COLLECTOR/DRAIN/ANODE of the test fixture.
Section
3.3.9
④
⑤
EMITTER/SOURCE/CATHODE
SENSE
Connect it with the terminal that corresponds to
EMITTER/SOURCE/CATHODE of the test fixture.
Section
3.3.9
Table 2.20 Name and function of each part on test adapter CS-508 (SMD CHIP type adapter)
Содержание CS-3000 Series
Страница 1: ...Instruction Manual Curve Tracer CS 3000 series CS 3100 CS 3200 CS 3300 ...
Страница 2: ...Ⓒ 2010 IWATSU ELECTRIC CO LTD All rights reserved ...
Страница 16: ......
Страница 170: ......
Страница 210: ...Chapter 4 Characteristics Measurement 4 40 Curve Tracer CS 3000 Series Instruction Manual Memo ...
Страница 214: ...Chapter 5 Daily Maintenance and Calibration 5 4 Curve Tracer CS 3000 Series Instruction Manual Memo ...
Страница 221: ...Chapter 6 Specifications Curve Tracer CS 3000 Series Instruction Manual 6 7 6 3 3 Test fixture CS 302 Unit mm ...
Страница 223: ...Chapter 6 Specifications Curve Tracer CS 3000 Series Instruction Manual 6 9 6 3 5 Test fixture CS 301 190 Unit mm ...