Capter 3
Function Setting and Basic Operation
3-16
Curve Tracer CS-3000 Series Instruction Manual
3.3.10 Connecting the measured device with the test adapter CS-510 (Option)
[Connection procedure]
(1) Open the cover of the test fixture.
If the cover is opened, this instrument is changed to INTERLOCK state and its output is shut
off.
(2) Connect the test adapter CS-510 with the measured device.
Insert the terminals of the measured device into the connection grooves of the socket.
- For CS-510, mount the device to be measured of following package SC-62 type on the socket.
(3) Mount the test adapter on the test fixture.
When mounting the test adapter, push it into the text fixture vertically while keeping its rear
surface parallel to the mounting part of the patch panel.
(4) When the cover of the test fixture is closed after connecting the test adapter, INTERLOCK state
is released and measurement can be carried out.