Minifors 2 - Benchtop Bioreactor
Accessories
07 February 2019
Page
53
of
214
5.11 Exit Gas Cooler
The basic unit supplies cooling liquid to the exit gas cooler. The
cooling liquid flow rate can be adjusted using the control valve on
the basic unit. The two hoses for water supply (bottom) and return
(top) are connected to the basic unit at the factory and simply con-
nected to the exit gas cooler via the two rapid couplings.
A piece of pressure hose is fitted onto the exit gas cooling pipe and
equipped with an exit gas filter. The hose connections and the exit
gas filter are secured with hose clamps. The hose clamp securing
the exit gas filter is equipped with a screw that can be used to
loosen or tight it manually.
1 Pressure hose und hose clamp
2 Exit gas filter
3 Hose connections:
a) Water outlet
b) Water inlet
4 Hoses with rapid couplings for water supply and return
(connected before delivery!)
5 Screw thread
The exit gas cooler is fitted with an O-ring before installation.
A screw thread is used to mount it in the 12 mm/Pg13.5 vessel top
plate port.
INFORMATION
The exit gas filter must be replaced with a new filter after each
cultivation process.
The exit gas cooler only works when the temperature control
system is switched on.
1
2
a)
3
b)
4
5