12 Interferences
The following section describes possible reasons for interferences
and how to resolve them. Reduce the service intervals in corre-
spondence with the actual loads if interferences become increas-
ingly common. Contact the manufacturer for interferences that can-
not be resolved by following the above instructions. For service
contact details, see page 2.
12.1 Interferences Basic Unit and Operating Panel
Interference
Equipment does not work
Possible cause
Remedy
By
Power cable not connected.
Connect the power cable.
Operator
Main switch not switched on.
Switch on the main switch.
Operator
Fuses blown.
Check the fuses of the mains connection and the
basic unit.
Qualified elec-
trician
Power connection interrupted.
Check the plug and cable at the basic unit and
mains socket.
Check mains cable for damage and kinks.
Qualified elec-
trician
LED flashes red, Equipment Alarm is
shown on the display, power failure
during a running Batch (process).
Acknowledge the alarm message, start the batch
again if necessary.
Operator
LED flashes red, Equipment Alarm is
shown on the display, control system
communication is interrupted.
Acknowledge the alarm message, contact the manu-
facturer’s service department if the fault message re-
turns.
Operator,
INFORS HT
service techni-
cian or spe-
cialist dealer
LED flashes red, Equipment Alarm is
shown on the display, pressure in
the culture vessel is too high.
Acknowledge the alarm message, if necessary re-
place the exit gas filter or reduce the gas flow rate.
Operator