Minifors 2 - Benchtop Bioreactor
Technical Data
07 February 2019
Page
211
of
214
14.5.5 Gassing
All gas enters via the sparger.
The specific gas flow rate, calculated for the maximum working vol-
ume, is 2 min
-1
for every vessel size.
Variants
Gas(ses)
Controlling flow rate
MFC accuracy
Air
2 MFC
1
(mass flow controller)
± 0.05 slm
Air + O
2
Air + N
2
1
).. 2 are preinstalled even for gassing using air only
Vessel
Control range of gas flow
l min
-1
TV
2
Max. AV
3
1.5 L
1.0 L
0.05 to 2
3.0 L
2.0 L
0.05 to 4
6.0 L
4.0 L
0.05 to 8
2
) Total volume
3
) Working volume
14.5.6 pH
Control system
Pumps
Acid
and
base
Measurement system, digital
Conventional pH sensor (potential measurement against refer-
ence) with built-in electronics
Type
Easyferm Plus ARC
Manufacturer
HAMILTON
Measuring range
pH 2 to pH 14
INFORMATION
The pH sensors are preconfigured by unit manufacturer IN-
FORS HT. Replacement sensors must be configured before
use.