Minifors 2 - Benchtop Bioreactor
Operation
Page
174
of
214
07 February 2019
6.
Enter the setpoint 0 (%) in the input field.
The input field turns orange and displays the set setpoint.
A dialogue appears with the notice
Make sure that N2 is con-
nected to the Gas2 connector
Make sure that N2 is connected to the correct connection.
7.
If necessary, connect the nitrogen and confirm with
OK
.
Gassing with nitrogen is activated, the stirrer switches on sim-
ultaneously.
8.
Wait until the oxygen has been displaced from the medium,
i.e. wait until the measured value (display via CALIBRATE
POINT 1) is stable.
9.
Press CALIBRATE POINT 1 to begin calibration.
Gassing and the stirrer are stopped.
CONFIRM NOW
creates a visual representation of the calibra-
tion procedure and slowly turns orange.
The procedure takes 3 minutes to guarantee a stable value.
When the countdown has finished, CONFIRM NOW is com-
pletely coloured in.
INFORMATION
If the measured value is assumed to be already stable, the wait-
ing time can be skipped by pressing CONFIRM NOW to con-
tinue with the second calibration point.
In the event of a problem, an error message is displayed with
a corresponding notice.