Minifors 2 - Benchtop Bioreactor
Technical Data
Page
210
of
214
07 February 2019
Accuracy
Measuring:
± 5 min
-1
at 100 min
-1
to 1,600 min
-1
1 % setpoint at > 100 % min-1
Control:
1 % full scale
Direction of stirrer
shaft’s rotation
Anti-clockwise = to left
Bearing
Outside vessel in drive hub
1)
Rotation speed is valid for 2 impellers, viscosity similar to water, with-
out gassing.
Impellers
Quantity / type
Material
Vessel
NW 145 and NW 115
Vessel size
NW 90
2 Rushton impellers
with 6 blades
Stainless steel, 316L,
electrolytically polished
PEEK
Vessel
A
B
C
6 l TV / NW 145
54 mm
11 mm
11 mm
3,0 l TV NW 115
46 mm
11 mm
11 mm
1.5 L TV/NW 90
38 mm
9 mm
11 mm
14.5.4 Temperature
Description
Value
Heating
Electrical, thermal block 630 W
Cooling
Tap water
1
via thermal block and
adapter
Sensor
Type: Pt100 1/3 DIN-B
Measuring range
0 °C to +145 °C
Control range
From 5 °C higher than preheating tem-
perature to 60 °C
Accuracy at +20 °C to
+60 °C
Measuring: ± 0.1 °C
Measuring: ± 0.2 °C
1)
A circulating cooler can be used instead of tap water for the cooling
system
.