Minifors 2 - Benchtop Bioreactor
Setup and Function
07 February 2019
Page
39
of
214
For more details on the electrode holder, see the “Accessories”
section, sub-section “Electrode Holder”.
3.9 Antifoam Control
Foam hinders the exchange of gas between the medium and the
gas phase in the head space. The exit gas filter can become
clogged with foam, which causes a pressure build-up in the vessel.
This can be prevented by adding antifoam agent.
The antifoam agent is kept in a reagent bottle that is connected to
the antifoam sensor and the antifoam pump via a hose. The sensor
also acts as a dosing needle. When the sensor comes in contact
with foam, the antifoam pump is activated and antifoam agent is
fed into the vessel via the dosing needle.
3.9.1 Antifoam Sensor
Inside diameter
2 mm
Hose connection outside diame-
ter
4 mm
A clamping adapter with a fixed O-ring is used to mount the sensor
in the 10 mm port in the vessel top plate.
1 Sensor head with port for banana connector (A)
2 Clamping adapter with slotted-head screw (B)
3 Needle with transparent insulation
The antifoam sensor is equipped with two NON-autoclavable pro-
tective caps.
2
1
A
B
3