Minifors 2 - Benchtop Bioreactor
Before Cultivation
07 February 2019
Page
75
of
214
Procedure
1.
Check the O-ring for sealing the top plate for damage and to
ensure that it is positioned correctly in the groove on the ves-
sel flange.
2.
Ensure that every component part is equipped with an intact
O-ring:
Check that the O-rings are correctly positioned and are un-
damaged. If necessary, reposition or replace. If component
parts are fitted into other component parts (clamping adap-
tor), there must also be an O-ring between them.
INFORMATION
Septum collars are sealed with a septum. No O-ring is used!
8.1.2 Mounting the Impellers
To mount the impellers to the stirrer shaft, proceed as follows:
Procedure
1.
Slide the impeller onto the stirrer shaft.
2.
Set the desired height.
3.
Tighten the grub screws on the impeller with the Allen key.
INFORMATION
To avoid unnecessary foam formation, do not fit the impeller at
the same height as the surface of the medium.