Minifors 2 - Benchtop Bioreactor
Operation
07 February 2019
Page
161
of
214
Header: Date of the last calibration
2-POINT CALIBRATION
: Select 2-point calibration
PRODUCT CALIBRATION
: Select product calibration (for
details, see the “pH product calibration” section).
SHOW SENSOR STATUS
: Shows the data and values
produced by the firmware of the sensor manufacturer that is
integrated in the sensor.
4.
Select 2-point calibration.
The menu display changes and displays the following:
Left-hand side
First Calibration Point
: Display of current calibration
point (first calibration point)
1
Immerse pH sensor into the first buffer
: Hold the pH
sensor in the buffer solution of the first calibration point
2
Select the pH of the first calibration buffer
: Select the
pH of the buffer solution for the first calibration point.