8.1.6 Moistening/Filling the Culture Vessel
If the medium in the culture vessel is to be autoclaved, this can
take place before the top plate is put in position and the additional
component parts are mounted.
Note the following about filling the culture vessel before autoclav-
ing:
Before autoclaving, only top up with heat-resistant media.
During autoclaving, evaporation may result in a loss of volume
and thus to increased salt concentration in the medium. If nec-
essary, top up with sterile water.
INFORMATION
Development of steam is not possible when autoclaving an
empty and dry culture vessel. Successful sterilisation is not
guaranteed.
Ensure that there is liquid in the culture vessel (approx. 10 mL of
water per litre of total volume).
8.1.7 Fitting the Vessel Top Plate
Proceed as follows to fit and fix the vessel top plate:
Procedure
1.
Fit the O-ring for the top plate gasket into the groove on the
edge of the vessel.
2.
Place the top plate carefully and with the correct alignment
into position and ensure that component parts do not touch
the baffles.