Minifors 2 - Benchtop Bioreactor
Operation
07 February 2019
Page
165
of
214
10.8.3 pH Sensor Product Calibration
It is possible to adjust the calibration curve to the current process
conditions using product calibration. This could be necessary if
there is a possibility of drift of the displayed pH during a long-term
cultivation, for example.
INFORMATION
Product calibration can only be carried out and is only effective if
the externally measured and entered pH value does not deviate
from the original pH value by more than 2 pH units.
Proceed as follows for a product calibration:
Procedure
1.
Call up the calibration menu of the pH parameter and wait for
the short initialisation phase to complete.
INFORMATION
The various menu displays of the initial calibration menu are not
displayed in detail in this section, they are shown in “Calibrating
the pH Sensor - Procedure”.
2.
In the menu display that follows the initialisation phase, press
PRODUCT CALIBRATION
The menu display changes and now displays the following: