A
Chamber rack with temperature probe
B
Temperature probe
Single-Base Extension and Stain
This step uses a robot to process the BeadChips.
CAUTION
The following steps must be performed without interruption.
1
At the robot PC, select XStain Tasks | XStain HD BeadChip.
2
If imaging immediately after staining, turn on the scanner to allow the lasers to stabilize.
3
Place reservoirs on the robot deck according to the deck map, and add reagents to reservoirs as follows:
Reagent
# BeadChips
Volume
95% formamide/1 mM EDTA
1–8
15 ml
9–16
17 ml
17–24
25 ml
RA1
1–8
10 ml
9–16
20 ml
17–24
30 ml
XC3
1–8
50 ml
9–16
100 ml
17–24
150 ml
4
Invert the XC1, XC2, TEM, STM, and ATM tubes to mix. Remove the caps, and place on the robot deck
according to the deck map.
Document # 11322427 v03
For Research Use Only. Not for use in diagnostic procedures.
62
Infinium HD Super Assay Reference Guide