Preparation
1
Remove each hybridization chamber from the hybridization oven. Allow to cool for 30 minutes before
opening.
2
Prepare the following items:
u
Fill two wash dishes with 200 ml PB1 each and label them Wash 1 and Wash 2.
u
Using a graduated cylinder, fill the Multi-Sample BeadChip Alignment Fixture with 150 ml PB1.
3
Remove the following Te-Flow flow-through chamber components from storage:
u
Black frames
u
Spacers (separated for ease of handling)
u
Clean glass back plates
u
Clamps
4
Prepare the following consumables for the subsequent
Extend and Stain BeadChips
step:
Item
Storage
Instructions
95% Formamide/1mM
EDTA
-25°C to -15°C
Thaw at room temperature. Invert 10 times to mix.
ATM
-25°C to -15°C
Thaw at room temperature. Invert 10 times to mix.
TEM
-25°C to -15°C
Thaw at room temperature. Invert 10 times to mix.
PB1
Room
temperature
Thaw at room temperature. Invert 10 times to mix.
RA1
-25°C to -15°C
Shake vigorously to resuspend. If necessary, vortex until
dissolved.
STM
-25°C to -15°C
Thaw at room temperature. Invert 10 times to mix.
XC1
-25°C to -15°C
Thaw at room temperature. Invert 10 times to mix.
XC2
-25°C to -15°C
Thaw at room temperature. Invert 10 times to mix.
XC3
Room
temperature
Thaw at room temperature. Invert 10 times to mix.
XC4
Room
temperature
Thaw at room temperature. Invert 10 times to mix.
Procedure
Wash BeadChips
1
Attach the wire handle and submerge the wash rack in Wash 1 containing 200 ml PB1.
NOTE
Replace PB1 in Wash 1 after every 12 BeadChips.
Document # 11322427 v03
For Research Use Only. Not for use in diagnostic procedures.
54