A
Chamber rack
B
Water circulator with programmable temperature controls
C
Reservoir cover
3
Confirm the actual temperature using the chamber rack temperature probe.
The temperature displayed on the water circulator screen can differ from the chamber rack temperature.
4
Remove bubbles trapped in the chamber rack.
a
Separate the heat exchanger from the reagent pan.
b
Lift the heat exchanger upright and away from you with the tubing at the bottom, and turn 90°
counter clockwise.
c
Return the heat exchanger to a horizontal position.
d
Repeat steps
and
3 times for a total of 4 rotations or until all bubbles are removed.
e
Using Kimwipes dampened with laboratory-grade water, clean all surfaces between the heat
exchanger and reagent pan. Discard Kimwipes with formamide waste.
f
Place the Te-Flow back on the reagent pan. Using the two guide pins in the reagent pan, make sure
that the Te-Flow is flush.
5
Using the Illumina temperature probe, test at least three locations on the chamber rack.
a
For accurate measurements, make sure that the temperature probe touches the base of the
chamber rack.
b
Make sure that all locations are at 44°C ± 0.5°C.
c
If the temperature is not within ± 0.5°C, adjust the water circulator control knob to reach 44°C ±
0.5°C.
d
Do not leave the temperature probe in the first three rows of the chamber rack. Reserve these rows
for BeadChips.
Document # 11322427 v03
For Research Use Only. Not for use in diagnostic procedures.
61
Infinium HD Super Assay Reference Guide