Figure 5
Robot Setup for Fragment MSA1
A
FMS
B
MSA1 Plate
4
Place FMS tubes in the robot tube rack according to the bed map. Remove the cap mat.
5
At the robot PC, select Run.
6
When the process completes, select OK.
7
Remove the plate from the robot bed and seal with a cap mat.
8
Vortex at 1600 rpm for 1 minute.
9
Pulse centrifuge at 280 × g.
10 Place the sealed plate on the 37°C heat block for 1 hour.
If you are continuing, leave the plate in the 37°C heat block until you have completed preparation for the
next step. Do not leave the plate on the heat block for longer than 2 hours.
SAFE STOPPING POINT
If you are stopping, seal the plate(s), and store at -25°C to -15°C.
Precipitate DNA
This step uses 100% 2-propanol and PM1 to precipitate the DNA.
Consumables
u
100% 2-propanol
u
PM1(2 tubes)
u
Cap mat
Preparation
1
Do one of the following:
Document # 11322427 v03
For Research Use Only. Not for use in diagnostic procedures.
42