A
Reservoir at barcode end of glass back plate
B
Glass back plate in position
8
Secure each flow-through chamber assembly with metal clamps as follows.
a
Using one finger, gently push the glass back plate against the alignment bar.
b
Place a metal clamp around the flow-through chamber 5 mm from the top edge.
c
Place a second metal clamp around the flow-through chamber at the barcode end, 5 mm from the
bottom of the reagent reservoir.
A
One stripe is visible between the first clamp and the alignment bar.
B
Glass back plate pressed against alignment bar.
C
Stripes are not visible between the second clamp and the barcode.
9
Remove the assembled flow-through chamber from the alignment fixture.
10 Starting at the nonbarcode end, trim the spacers from each end of the assembly using scissors.
Document # 11322427 v03