5
Place a
clear
spacer onto the top of each BeadChip. Use the alignment fixture grooves to guide the
spacers into position.
The white spacers are not a substitute for the clear spacers.
6
Place the alignment bar onto the alignment fixture. Fit the groove on the alignment bar over the tab on the
alignment fixture.
7
Place a clean glass back plate on top of each clear spacer. Position the plate reservoir at the barcode
end of the BeadChip, facing inward to create a reservoir against the BeadChip surface.
A
Reservoir at barcode end of glass back plate
B
Glass back plate in position
8
Secure each flow-through chamber assembly with metal clamps as follows.
a
Using one finger, gently push the glass back plate against the alignment bar.
b
Place a metal clamp around the flow-through chamber 5 mm from the top edge.
c
Place a second metal clamp around the flow-through chamber at the barcode end, 5 mm from the
bottom of the reagent reservoir.