u
Regularly check lid-base pairs to make sure that they fit securely. Also check hinges for signs of abnormal
wear or loose fittings. An airtight seal requires that the hinges have adequate clamping strength.
u
Record which hybridization chamber was used for each BeadChip. If sample evaporation or other
processing anomalies occur, investigate the appropriate hybridization chambers.
u
When the hybridization chamber inserts contain BeadChips, keep them steady and level when lifting or
moving.
u
Avoid shaking, and always keep parallel to the lab bench.
u
Do not hold by the sides near the sample inlets.
Rocker Platform [Optional]
If you are using an Illumina Hybridization Oven with a rocker platform, follow these guidelines:
u
If 7 to12 hybridization chambers (Hyb Chambers) are placed in 1 hybridization oven, the rocker must be
removed before beginning the
Hybridize to BeadChip
section.
u
If 1 to 6 Hyb Chambers are placed in 1 hybridization oven, removing the rocker platform before beginning
the
Hybridize to BeadChip
section is optional.
u
Removing the rocker platform is optional for plate incubation in the
Amplify DNA, Incubate DNA,
Fragment DNA, Precipitate DNA,
and
Resuspend DNA
sections.
Acronyms
Acronym
Definition
EDTA
Ethylenediaminetetraacetic acid
EtOH
Ethanol
ATM
Anti-Stain Two-Color Master Mix
FMS
Fragmentation solution
MA1
Multi-Sample Amplification 1 Mix
MA2
Multi-Sample Amplification 2 Mix
MSM
Multi-Sample Amplification Master Mix
PB1
Reagent used to prepare BeadChips for hybridization
PB2
Humidifying buffer used during hybridization
PM1
Precipitation solution
RA1
Resuspension, hybridization, and wash solution
STM
Superior Two-Color Master Mix
TEM
Two-Color Extension Master Mix
XC1
XStain BeadChip solution 1
XC2
XStain BeadChip solution 2
XC3
XStain BeadChip solution 3
XC4
XStain BeadChip solution 4
Illumina LIMS
If you are using the automated protocol with Illumina LIMS, follow these guidelines:
u
At each step in the protocol, make sure that the Use Barcodes checkbox is selected.
Document # 11322427 v03
For Research Use Only. Not for use in diagnostic procedures.
5
Infinium HD Super Assay Reference Guide