4
Immediately cover the chamber with the lid to prevent evaporation.
5
Leave the closed chambers on the benchtop at room temperature until the BeadChips are loaded with
DNA (~1 hour).
6
[Illumina LIMS]
Select Infinium HD Super | Confirm for Hyb.
a
Scan the barcode of each MSA1 plate you plan to hybridize.
b
Scan the BeadChip barcode on the package of each BeadChip you plan to hybridize.
c
Select Verify.
Prepare the Robot
1
Remove all BeadChips from packaging.
2
Place BeadChips into the robot BeadChip alignment fixtures. Align the barcode end with the ridges
stamped into the robot BeadChip alignment fixture.
3
Stack the robot BeadChip alignment fixtures and carry them to the robot.
4
Place the robot BeadChip alignment fixtures onto the robot deck according to the deck map in
5
Pulse centrifuge the MSA1 plate at 280 × g.
6
Place the MSA1 plate onto the robot deck according to the deck map, and remove the heat seal.
Document # 11322427 v03
For Research Use Only. Not for use in diagnostic procedures.
49