Figure 8
Robot Deck Setup for Hybridization
*
A
MSA1 Plate
B
Robot BeadChip Alignment Fixtures
CAUTION
BeadChips must be transferred to hybridization chambers immediately at the end of the next
procedure
(Start the Robot)
. Do not begin
Start the Robot
if you cannot immediately transfer the
BeadChips.
Start the Robot
1
At the robot PC, select Run.
u
The robot scans the barcode on the BeadChips to confirm the correct BeadChips are loaded.
2
Place each robot tip alignment guide on top of each robot BeadChip alignment fixture.
a
Make sure that the Guide-E barcode is upside down and facing away from you.
b
Push both the tip guide and alignment fixture to the upper left corner in its section of the robot bed.
3
At the robot PC, select OK.
u
The robot scans the barcode on the robot tip alignment guide to confirm that the correct tip guide is
being used.
u
The robot dispenses DNA sample to the BeadChips.
u
When the process is complete, the robot PC sounds an alert and opens a message box.
Document # 11322427 v03
For Research Use Only. Not for use in diagnostic procedures.
50
Infinium HD Super Assay Reference Guide