Figure 6
Guide E Robot Tip Alignment Guide
2
Wash and dry the robot tip alignment guide. For washing instructions, see
.
Procedure
Denature DNA
1
Place the MSA1 plate on the preheated heat block for 20 minutes to denature the DNA.
2
Cool the MSA1 plate on the benchtop at room temperature for 30 minutes.
3
Pulse centrifuge at 280 × g.
Document # 11322427 v03
For Research Use Only. Not for use in diagnostic procedures.
47
Infinium HD Super Assay Reference Guide