2
Remove the hybridization chamber inserts from the hybridization chambers.
3
Remove the BeadChips from the hybridization inserts.
4
Remove the cover seals from the BeadChips.
Using powder-free gloved hands, hold the BeadChip securely and by the edges in one hand. Remove
the entire seal in a single, continuous motion. Do not touch exposed arrays.
5
Immediately and carefully slide each BeadChip into the wash rack in Wash 1, making sure that the
BeadChip is submerged in PB1 a maximum of 8 BeadChips)..
6
Repeat steps
–
until all BeadChips are transferred to the submerged wash rack in Wash 1.
Document # 11322427 v03
For Research Use Only. Not for use in diagnostic procedures.
55
Infinium HD Super Assay Reference Guide